Determination of Diagnostic Reference Level (DRL) in Non-Contrast and Contrast Thorax CT Scan Examinations at Bali Mandara Regional Hospital

Authors

Ni Luh Emy Pramitha , Gusti Ngurah Sutapa , Ni Luh Putu Trisnawati , Ni Nyoman Ratini , Hery Suyanto , Winardi Tjahyo Baskoro

DOI:

10.29303/jppipa.v11i7.10344

Published:

2025-07-25

Downloads

Abstract

A study has been conducted to determine the Diagnostic Reference Level (DRL) in non-contrast and contrast chest CT scan examinations at Bali Mandara Hospital. The study was conducted at the Radiology Installation of Bali Mandara Hospital using dose reports from non-contrast and contrast chest CT Scan examinations with data obtained including age, gender, kV, mAs, slice thick, time rotation, Pitch, as patient information and protocols used in the examination and dose information in the form of CTDIvol and DLP. This study aims to determine the local DRL value against the national DRL value. The method used to determine the DRL value in this study is by collecting data in the form of recording and documenting the dose report of CT Scan examination patients, with the data for each examination being as many as 30. From the recorded data, the CTDIvol and DLP values are used to find the DRL value by sorting it from the smallest to the largest value and then calculating the 3rd quartile as a determinant of the local DRL. The interpolation method is also used to consider the position of the 3rd quartile so that it is at a decimal value. From the research results, the CTDIvol value for non-contrast thorax examination was 6.43 mGy, and the DLP value was 293.08 mGy.cm. Meanwhile, in contrast to the thorax examination, the CTDIvol value was 6.14 mGy, and the DLP value was 852.57 mGy.cm. Furthermore, a statistical analysis was carried out in the form of a one-way t-test using SPSS software, which was used to compare the DRL values of both CTDIvol and DLP non-contrast and contrast thorax examinations with the standard values set by BAPETEN 2021 as a reference used in Indonesia. Based on the results of the analysis with a one-way t-test, it can be seen that the DRL values for non-contrast thorax examinations (DRL CTDIvol and DLP) and contrast examinations (DRL CTDIvol) do not exceed the standards set by BAPETEN. However, in the contrast thorax CT Scan examination, the DRL DLP value exceeded the standard value set by BAPETEN, so dose optimization was needed by re-evaluating the protocol implemented at Bali Mandara Hospital.

Keywords:

CT scan DRL Thorax contras Thorax non contras

References

Abimanyu, B., Rusyadi, L., & Taufiq, T. (2017). Analisis Informasi Citra Anatomi Fase Late Artery dengan Variasi Time Scan Delay pada Pemeriksaan MSCT Abdomen. Jurnal Imejing Diagnostik (JImeD), 3(1), 213–219. https://doi.org/10.31983/jimed.v3i1.3187

Anam, C., Mahdani, F. R., Dewi, W. K., Sutanto, H., Triadyaksa, P., Haryanto, F., & Dougherty, G. (2021). An improved method for automated calculation of the water‐equivalent diameter for estimating size‐specific dose in CT. Journal of Applied Clinical Medical Physics, 22(9), 313–323. https://doi.org/10.1002/acm2.13367

BAPETEN. (2020). Peraturan Badan Pengawas Tenaga Nuklir Republik Indonesia Nomor 4 Tahun 2020 Tentang Keselamatan Radiasi Pada Penggunaan Pesawat Sinar-X Dalam Radiologi Diagnostik Dan Intervensional. Peraturan Badan Pengawas Tenaga Nuklir Republik Indonesia. Retrieved from https://jdih.bapeten.go.id/unggah/dokumen/peraturan/1028-full.pdf

Bushberg, J. T. (2012). The essential physics of medical imaging. Wolters Kluwer Health/Lippincott Williams & Wilkins.

Chau, S., & Hayre, C. M. (2022). Computed Tomography: A Primer for Radiographers. CRC Press. https://doi.org/10.1201/9781003132554

Chuninghum, J. (1983). The Physics of Radiology (Fourth). Charles C Thomas Publiser.

Coakley, F. V, Gould, R., Yeh, B. M., & Arenson, R. L. (2011). CT Radiation Dose: What Can You Do Right Now in Your Practice? American Journal of Roentgenology, 196(3), 619–625. https://doi.org/10.2214/AJR.10.5043

Dewi, R., Jeniyanthi, N. P. R., & Kristin, P. M. (2025). Analisis Pemeriksaan CT Scan Thorax dengan Kontras pada Klinis Tumor Paru di Rs Bhayangkara Makassar. Jurnal Sehat Indonesia (JUSINDO), 7(1), 439–451. https://doi.org/10.59141/jsi.v7i01.211

Diartama, A. A. A., Suputra, C. G. B., Mahendrayana, I. M. A., & Wirajaya, I. W. A. (2024). Evaluasi Nilai CTDIvol dan DLP pada pemeriksaan CT Scan Stonografi Periode Januari Sampai Maret 2023 di RSUD Sanjiwani Gianyar. Jurnal Medika Malahayati, 8(1), 89–95. https://doi.org/10.33024/jmm.v8i1.12721

Dinayawati, D., Sunoto, J., & Handis, L. A. (2024). Penerapan Optimisasi Paparan Medik CT-Scan Melalui Tingkat Panduan Diagnostik di RSUP dr. Soeradji Tirtonegoro. MJS Medical Journal of Soeradji, 1(1), 20–33. https://doi.org/10.70605/t6r1wj90

Duandini, E., Etika, E. A., Nurulita, S. F., & Hidayanto, E. (2021). Analisis Perbandingan Diagnostic Reference Level (Drl) Modalitas Ct Scan Sebagai Upaya Optimasi Proteksi Dan Keselamatan Radiasi Di Berbagai Negara. Berkala Fisika, 24(3), 100–108. Retrieved from https://ejournal.undip.ac.id/index.php/berkala_fisika/article/view/44183

Ginting, V. S. B., Sutapa, G. N., Ratnawati, I. G. A. A., & Widiana, I. K. (2023). Analisis Diagnostic Reference Level (DRL) Pasien Pada Pemeriksaan Abdomen Kontras dan Nonkontras Dengan Pesawat CT Scan Di RSUD Bali Mandara. Kappa Journal, 7(1), 165–169. https://doi.org/10.29408/kpj.v7i1.18985

Hussain, S., Mubeen, I., Ullah, N., Shah, S. S. U. D., Khan, B. A., Zahoor, M., Ullah, R., Khan, F. A., & Sultan, M. A. (2022). Modern Diagnostic Imaging Technique Applications and Risk Factors in the Medical Field: A Review. BioMed Research International, 2022(1). https://doi.org/10.1155/2022/5164970

Imai, R., Miyazaki, O., Horiuchi, T., Kurosawa, H., & Nosaka, S. (2015). Local diagnostic reference level based on size-specific dose estimates: Assessment of pediatric abdominal/pelvic computed tomography at a Japanese national children’s hospital. Pediatric Radiology, 45(3), 345–353. https://doi.org/10.1007/s00247-014-3189-4

Irsal, M., Targian, A., Elisa Wulandari, S., & Kinasih, N. (2023). Evaluasi Diagnostic Reference Levels (DRL) Pada Pemeriksaan Thorax PA di Pulau Jawa Berdasarkan Database DRL BAPETEN. Jurnal Pengawasan Tenaga Nuklir, 3(2), 88–96. https://doi.org/10.53862/jupeten.v3i2.014

Kathon, B. O., Hartoyo, P., & Samsun, S. (2022). Uji Resolusi Spasial dan Slice Thickness Pada CT Scan 128 Dan 16 Slice Dengan Menggunakan Phantom Quart DVT-AP. Jurnal Pembelajaran Fisika, 11(3), 123. https://doi.org/10.19184/jpf.v11i3.33956

Khusniatul, P., Hidayanto, E., Arifin, Z., & Anam, C. (2014). Pengaruh Variasi Faktor Eksposi (Tegangan Tabung Dan Arus Waktu) Serta Pitch Terhadap Computed Tomography Dose Index (Ctdi) Di Udara Menggunakan Ct Dose Profiler. Youngster Physics Journal, 3(4), 363–372. Retrieved from https://ejournal3.undip.ac.id/index.php/bfd/article/view/7078

Latifah, R., Jannah, N. Z., Nurdin, D. Z. ., & P, B. (2019). Determination of Local Diagnostic Reference Level (Ldl) Pediatric Patients on Ct Head Examination Based on Size-Specific Dose Estimates (Ssde) Values. Journal Of Vocational Health Studies, 2(3), 127. https://doi.org/10.20473/jvhs.v2.i3.2019.127-133

Lestari, L., & Susilo, S. (2022). Detection and Characterization of Lung Cancer using CT Scan Technology. Proceedings of the Soedirman International Conference on Mathematics and Applied Sciences, 5(2021), 91–94. https://doi.org/10.2991/apr.k.220503.019

Maldjian, P. D., & Goldman, A. R. (2013). Reducing Radiation Dose in Body CT: A Primer on Dose Metrics and Key CT Technical Parameters. American Journal of Roentgenology, 200(4), 741–747. https://doi.org/10.2214/AJR.12.9768

Mardliyyah, A., Sensusiati, A. D., & Sari, A. K. (2020). Role Of Radiographer In Handling Covid-19 At CT Scan Room During Pandemic. Journal of Vocational Health Studies, 4(2), 83. https://doi.org/10.20473/jvhs.V4.I2.2020.83-88

Missinychrista, R. A., Subagiada, K., & Putri, E. R. (2023). A Comparison of CT-Scan Output Doses and Doses Given to Kidney Stone Patients. Jurnal Fisika Flux: Jurnal Ilmiah Fisika FMIPA Universitas Lambung Mangkurat, 20(3), 223. https://doi.org/10.20527/flux.v20i3.17098

Noveranty, A., Purwaningsih, S., & Fendriani, Y. (2024). Analisis Pengaruh Variasi Faktor Eksposi Pada Ct Scan Terhadap Kualitas Citra Dan Dosis Radiasi Pada Pemeriksaan Abdomen. Journal Online Of Physics, 9(3), 53–59. https://doi.org/10.22437/jop.v9i3.35155

Osman, N. D., Shamsuri, S. B. M., Tan, Y. W., Razali, M. A. S. M., & Isa, S. M. (2017). A single institution study of radiation dose received from CT imaging: A comparison to Malaysian NDRL. Journal of Physics: Conference Series, 851(1), 012003. https://doi.org/10.1088/1742-6596/851/1/012003

Paulo, G., Damilakis, J., Tsapaki, V., Schegerer, A. A., Repussard, J., Jaschke, W., Frija, G., Hierath, M., & Clark, J. (2020). Diagnostic Reference Levels based on clinical indications in computed tomography: a literature review. Insights into Imaging, 11(1), 96. https://doi.org/10.1186/s13244-020-00899-y

Pomara, C., Pascale, N., Maglietta, F., Neri, M., Riezzo, I., & Turillazzi, E. (2015). Use of contrast media in diagnostic imaging: medico-legal considerations. La Radiologia Medica, 120(9), 802–809. https://doi.org/10.1007/s11547-015-0549-6

Pratama, I. B. G. P., & Rusmanto. (2020). Establishment Diagnostic Reference Level for CT-Scan Procedure in Indonesia. Journal of Physics: Conference Series, 1505(1), 012037. https://doi.org/10.1088/1742-6596/1505/1/012037

Puspita, M. I., Utama, H. N., & Felayani, F. (2018). Eknik Pemeriksaan Computed Tomography Scanning (Ct-Scan) Thoraks Dengan Kasus Massa Pulmo Di Instalasi Radiologi Rspau. Dr. S. Hardjolukito. Jurnal DIII Teknik Rontgen, 4(2), 9–13. Retrieved from http://stikeswh.ac.id:8082/journal/index.php/jitk/article/view/86/80

Raharja, H. D. R., Hariyati, I., N. Haposan, U., Prawoto, U., R. Fajrin, D., R. Y. Hanapi, M., I. S. H. Supit, N., Salamah, T., & E. Lubis, L. (2023). Audit Dosis dan Optimisasi Pemeriksaan CT Scan Kepala: Studi Multisenter. Jurnal Pengawasan Tenaga Nuklir, 3(2), 79–87. https://doi.org/10.53862/jupeten.v3i2.013

Sari, A. W., & Fransiska, E. (2018). Pengaruh Faktor Eksposi dengan Ketebalan Objek pada Pemeriksaan Foto Thorax Terhadap Gambaran Radiografi. Journal of Health, 5(1), 17. https://doi.org/10.30590/vol5-no1-p17-21

Seeram, E., & Sil, J. (2013). Computed tomography: Physical principles, instrumentation, and quality control. Practical SPECT/CT in Nuclear Medicine, 77–107. https://doi.org/10.1007/978-1-4471-4703-9_5

Silvia, H., Milvita, D., Prasetio, H., & Yuliati, H. (2013). Estimasi Nilai Ctdi Dan Dosis Efektif Pasien Bagian Head. Jurnal Fisika Unand, 2(2), 128–134. Retrieved from https://jfu.fmipa.unand.ac.id/index.php/jfu/article/view/35/35

Strong, A. T., Sharma, G., Tu, C., Aminian, A., Young, J. B., Rodriguez, J., & Kroh, M. (2018). A Population-Based Study of Early Postoperative Outcomes in Patients with Heart Failure Undergoing Bariatric Surgery. Obesity Surgery, 28(8), 2281–2288. https://doi.org/10.1007/s11695-018-3174-3

Sudirman, Z. D., Prasetya, I. M. L., & Mahendrayana, I. M. A. (2024). Analisis Pemeriksaan Msct Scan Thorax Dengan Kontras Media Intravena Pada Klinis Yang Terdiagnosis Tumor Paru Di Rs Bhayangkara Makassar. Prepotif : Jurnal Kesehatan Masyarakat, 8(3), 5161–5171. https://doi.org/10.31004/prepotif.v8i3.35036

Sumarsono, S., & Musdalifah, I. (2020). Hubungan Indeks Massa Tubuh Terhadap Dosis Radiasi Pada Angiografi Koroner. Lontara, 1(2), 108–116. https://doi.org/10.53861/lontarariset.v1i2.77

Susanto, W. (2018). Penentuan Nilai Diagnostic Reference Level (DRL) CT-Scan Untuk Pemeriksaan Kepala dan Dada Pasien Dewasa. Badan Pengawas Tenaga Nuklir (BAPETEN).

Wangko, L. C., Budiono, B., & Lefrandt, R. L. (2013). Angiografi Koroner Indikasi, Kontraindikasi, Dan Proteksi Terhadap Radiasi. Jurnal Biomedik (JBM), 4(3), 150–155. https://doi.org/10.35790/jbm.4.3.2012.794

Yogantara, G. A. K., Putu, Sutapa, N., Gusti, Yuliara, M., & I. (2021). Analisis Dosis Efektif Pada Pemeriksaan Computer Tomography (CT) Scan Kepala Di RSUD Sanjiwani Gianyar Effective Dose Analysis on Computer Tomography (CT) Head Scan at Gianyar Sanjiwani Hospital. Jurusan Fisika Fakultas Matematika Dan Ilmu Pengetahuan Alam, 22(2), 53–59. https://doi.org/10.24843/BF.2021.v22.i02.p01

Author Biographies

Ni Luh Emy Pramitha, Universitas Udayana

Author Origin : Indonesia

Gusti Ngurah Sutapa, Undayana University

Author Origin : Indonesia

Ni Luh Putu Trisnawati, Universitas Udayana

Author Origin : Indonesia

Ni Nyoman Ratini, Universitas Udayana

Author Origin : Indonesia

Hery Suyanto, Universitas Udayana

Author Origin : Indonesia

Winardi Tjahyo Baskoro, Universitas Udayana

Author Origin : Indonesia

Downloads

Download data is not yet available.

How to Cite

Pramitha, N. L. E., Sutapa, G. N., Trisnawati, N. L. P., Ratini, N. N., Suyanto, H., & Baskoro, W. T. (2025). Determination of Diagnostic Reference Level (DRL) in Non-Contrast and Contrast Thorax CT Scan Examinations at Bali Mandara Regional Hospital. Jurnal Penelitian Pendidikan IPA, 11(7), 884–893. https://doi.org/10.29303/jppipa.v11i7.10344